The Application of Step Filter in High Energy Photon Beams
고에너지 광자선 조사면내 계층필터의 응용
최태진; 김옥배
Abstract
The half-value layer (HVL) of cerrobend alloy [p~9.4g/ccj was examined by experimental measurements for 6MV X-ray beams. The output factors are a function of field size; thickness of attenuator; especially those factors for a blocked field are less than that for the collimater open field size. Measured dose distributions under attenuators; which are blocked with partially transmitted slab at whole field size; were presented. The expected dose distributions under partial attenuators agree well with measured data; which has shown that partial attenuator method can be applied in radiation treatment planning.